Rigol M300 configuring a flexible automated tracking mainframe, this modular acquisition engine records high-accuracy electrical signals inside production quality control stations.
- 300V terminal protection isolates internal components.
- 10MΩ standard input impedance minimizes loading.
- 1.000000A top direct current range registers.
- 140 dB common mode rejection isolates.
- 200.0000mV low voltage span reads microvolts.
In Stock
Rigol M300 Data Acquisition Switch System Specs
| Function | Range [2] |
| DC Voltage | 200.0000mV |
| 2.000000V | - |
| 20.00000V | - |
| 200.0000V | - |
| 300.000V | - |
| DC Current | 200.0000µA |
| 2.000000mA | <0.25V |
| 20.00000mA | <0.07V |
| 200.0000mA | <0.7V |
| 1.000000A | <0.12V |
| Resistance [4] | 200.0000Ω |
| 2.000000kΩ | 1mA |
| 20.00000kΩ | 100µA |
| 200.0000kΩ | 10µA |
| 1.000000MΩ | 2µA |
| 10.00000MΩ | 200nA |
| 200nA || 10MΩ | 0.300 + 0.010 |
| Input Impedance | 200mV, 2V, 20V ranges: 10MΩ or >10GΩ (For these ranges, input beyond ±26 V are clamped through 106 kΩ) |
| Input Protection | 300V |
| Input Offset Current | 50pA, at 25°C , typical |
| CMRR (common mode rejection ratio) | 140 dB for 1 kΩ unbalanced resistance in LO lead, ±300 VDC peak maximum. |
| Measurement Method | 4-wire or 2-wire resistance Current source referenced to LO input |
| Open-circuit Voltage | Limited to <10 V |
| Max. Lead Resistance (4-wire) | 10% of range per lead for 200 Ω and 2 kΩ ranges, 1 kΩ per lead on all other ranges |
| Input Protection | 300V on all ranges |
| Offset Compensation | Available on 200 Ω, 2 kΩ and 20 kΩ ranges. |
| Shunt Resistor | 100 Ω for 200 uA, 2 mA |
| Function | Range [2] |
| True RMS AC Voltage [4] | 200.0000mV |
| 5Hz-10Hz | 0.35 + 0.03 |
| 10Hz-20kHz | 0.04 + 0.03 |
| 20kHz-50kHz | 0.10 + 0.05 |
| 50kHz-100kHz | 0.55 + 0.08 |
| 100kHz- 300kHz | 4.00 + 0.50 |
| 2.000000V | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.02 |
| 10Hz-20kHz | 0.04 + 0.02 |
| 20kHz-50kHz | 0.10 + 0.04 |
| 50kHz-100kHz | 0.55 + 0.08 |
| 100kHz- 300kHz | 4.00 + 0.50 |
| 2.000000V | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.02 |
| 10Hz-20kHz | 0.04 + 0.02 |
| 20kHz-50kHz | 0.10 + 0.04 |
| 50kHz-100kHz | 0.55 + 0.08 |
| 100kHz-300kHz | 4.00 + 0.50 |
| 20.00000V | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.03 |
| 10Hz-20kHz | 0.04 + 0.04 |
| 20kHz- 50kHz | 0.10 + 0.05 |
| 50kHz-100kHz | 0.55 + 0.08 |
| 100kHz-300kHz | 4.00 + 0.50 |
| 200.0000V | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.02 |
| 10Hz-20kHz | 0.04 + 0.02 |
| 20kHz-50kHz | 0.10 + 0.04 |
| 50kHz-100kHz | 0.55 + 0.08 |
| 100kHz-300kHz | 4.0 + 0.50 |
| 300.000V | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.02 |
| 10Hz-20kHz | 0.04 + 0.02 |
| 20kHz-50kHz | 0.10 + 0.04 |
| 50kHz-100kHz | 0.55 + 0.08 |
| 100kHz-300kHz | 4.0 + 0.50 |
| True RMS AC Current [6] | 200.0000µA |
| 5Hz-10Hz | 0.35 + 0.06 |
| 10Hz-5kHz | 0.15 + 0.06 |
| 5kHz-10kHz | 0.35 + 0.70 |
| 2.000000mA | 3Hz-5Hz |
| 5Hz-10Hz | 0.30 + 0.04 |
| 10Hz-5kHz | 0.12 + 0.04 |
| 5kHz-10kHz | 0.20 + 0.25 |
| 20.00000mA | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.06 |
| 10Hz-5kHz | 0.15 + 0.06 |
| 5kHz-10kHz | 0.35 + 0.70 |
| 200.0000mA | 3Hz-5Hz |
| 5Hz-10Hz | 0.30 + 0.04 |
| 10Hz-5kHz | 0.10 + 0.04 |
| 5kHz-10kHz | 0.20 + 0.25 |
| 1.000000A | 3Hz-5Hz |
| 5Hz-10Hz | 0.35 + 0.06 |
| 10Hz-5kHz | 0.15 + 0.06 |
| 5kHz-10kHz | 0.35 + 0.70 |
| 5Hz-10Hz | 0.35 + 0.08 |
| 10Hz-5kHz | 0.15 + 0.08 |
| Measurement Method | AC-coupled True-RMS - measure the ac component of input with up to 300 V DC bias on any range. |
| Crest Factor | ≤ 5 at full range |
| Input Impedance | 1 MΩ ± 2%, in parallel with <150 pF capacitance on any range |
| Input Protection | 300 V rms on all ranges |
| AC Filter Bandwidth | Slow: 3 Hz - 300 kHz |
| CMRR (common mode rejection ratio) | 70 dB, for the 1 kΩ unbalance in LO lead, <60 Hz common mode signal frequency, ±500 VDC peak maximum. |
| Measurement Method | Direct coupled to the fuse and shunt; AC-coupled True RMS measurement (measure the AC component). |
| Crest Factor | ≤ 3 at full range |
| Max. Input | DC + AC current peak value <300% of range. Current with DC current component <10 A rms. |
| Shunt Resistor | 100 Ω for 200 uA, 2 mA |
| Function | Range |
| Frequency, Period | 200mV-300V |
| 5 Hz-10 Hz | 0.04 |
| 10 Hz-40 Hz | 0.02 |
| 40 Hz-300 kHz | 0.005 |
| 300 kHz-1 MHz | 0.005 |
| Frequency | Gate Time (Resolution) |
| 1s(0.1ppm) | 0.1s(1ppm) |
| 3 Hz-5Hz | 0 |
| 5 Hz-10Hz | 0 |
| 10 Hz-40Hz | 0 |
- Power Cord Conforming to the Standard of the Country
- USB Cable
- Mixed-interface Separator Line
- M300 Series Standard Control and Data Analysis PC Software
- Four Spare Fuses
- 2 AC, 250 V, T3.15 A fuses
- 2 AC, 250 V, T250 mA fuses
- Quick Guide
- NIST calibration certificate for the M300