Fluke 726 Multifunction Precision Process Calibrator Specs
| 30.000 V | 0.01%+ 2 mV (upper display) |
| 20.000 V | 0.01%+ 2 mV (lower display) |
| 90.00 mV | 0.01%+ 10µV |
| -10.00 mV to 75.00 mV | 0.025 % + 1 count (via TC connector) |
| 24.000 mA | 0.01%+ 2 mA |
| 0.0 to 4000.0 Ω | +/- 0.015% |
| 401 to 1500 Ω | 0.5Ω (4-wire), 1Ω (2- and 3-wire) |
| 1500 to 3200 Ω | 1Ω (4-wire), 1.5Ω (2- and 3-wire) |
| 2.0 to 1000.0 CPM | 0.05 % + 0.1 cpm |
| 1.0 to 1000.0 Hz | 0.05 % + 0.1 Hz |
| 1.00 to 15.00 kHz | 0.05 % + 0.01 kHz |
| Sensitivity | 1 V peak-to-peak-minimum |
| Accuracy | from 0.025% of range using any of 29 pressure modules.(for detailed specifications refer to pressure modules in options and accessories) Modules available for differential, gauge, vacuum, absolute, dual and high pressure. |
| 100.00 mV | 0.01%+10 µV |
| 20.000 V | 0.01%+2 mV |
| -10.00 mV to 75.00 mV | 0.025 % + 1 count (via TC connector) |
| 24.000 mA (Source) | 0.01%+ 2 µA |
| 24.000 mA (Simulate) | 0.01%+ 2 µA |
| Range, Accuracy (Percent of reading + floor) and Excitation Current | 5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5mA 5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3mA 400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA 1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA |
| 2.0 to 1000.0 CPM | 0.05% |
| 1.0 to 1000.0 Hz | 0.05% |
| 1.0 to 10.0 kHz | 0.25% |
| 10.0 to 15.0 kHz | 0.5% |
| Waveform: | 5 V p-p squarewave, -0.1 V offset |
| Cu10 NI-120 PT-100 (385) PT-100 (392) PT-100 (391-JIS) PT-200 (385) PT-500 (385) PT-1000 (385) RTD Resolution: Cu10.01 C J: K: T: E: R: S: B: L: U: N: XK: BP: TC Resolution | 1.8 °C 0.15 °C 0.45 °C 0.35 °C 0.35 °C 0.95 °C 0.55 °C 0.35 °C 0.01 C (0.01 F) 0.1 C (0.1 F) 0.4 °C 0.5 °C 0.4 °C 0.4 °C 1.4 °C 1.4 °C 1.4 °C 0.45 °C 0.45 °C 0.6 °C 0.4 °C 1.1 °C J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F |
| Source functions: | Voltage, current, resistance, frequency, temperature |
| Ramps | Slow ramp, Fast ramp, 25% step-ramp |
| Voltage: | 24 V |
| Accuracy: | 10% |
| Maximum current: | 22 mA, short circuit protected |
| Source functions: | Voltage, current, resistance, frequency, temperature |
| Steps | 25% of range, 100% of range |
| Operating Temperature | -10 °C to +55 °C |
| Storage Temperature | -20 °C to 71 °C |
| Operating Altitude | 3000 m |
| Agency Approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 |
| Size | 200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in) |
| Weight | 650 g (23 oz.) |
| Batteries | 4 AA alkaline batteries |
| Battery Replacement | Separate battery compartment, accessable without breaking calibration seal |
| Side port connections | Pressure module connector, also used for remote real-time programming |
- Fluke 726
- One set of AC72 alligator clips
- One set of stackable test leads
- One set of TL75 test leads
- 4 AA Batteries (installed)
- 725/726 CD-ROM (contains Users Manual)
- NIST- traceable calibration report and data