Additel ADT875PC-660-B-110V - Dry Well Calibrator Specs

Specification875-155
Temperature Range at 23℃-40°C to 155°C
Display Accuracy±0.18°C at Full Range
Stability (30 min)±0.01°C at Full Range
Axial Uniformity at 60 mm (2.4 in)±0.07°C at Full Range
±0.1°C at 200°C±0.3°C at 420°C
±0.2°C at 350°C±0.5°C at 660°C
Radial Uniformity±0.01°C at Full Range
±0.015°C at 200°C±0.05°C at 420°C
±0.02°C at 350°C±0.1°C at 660°C
Loading Effect±0.1°C (Display Sensor)
±0.02°C (External Sensor)±0.015°C (External Sensor)
Hysteresis (Display Sensor)0.025°C
Environmental Conditions8°C to 38°C guaranteed accuracy
Storage Conditions-20°C to 60°C
IP RatingIP20
Immersion Depth150 mm (5.9 in)
Insert OD25.8 mm (1.02 in)
Heating Time13 min: -40°C to 155°C
Cooling Time28 min: 155°C to -40°C
8 min: 155°C to 23°C10 min: 100°C to 50°C
20 min: 23°C to -40°C10 min: 50°C to 33°C
Typical Time to Stability10 min
Resolution0.01°C
Units°C, °F, and K
Display6.5 in (165 mm) color touch screen
Size (H x W x D)320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in)
Weight9.9 kg (21.8 lbs)
Power Requirements90-254 VAC, 45-65 Hz, 580 W
Mechanical TestingVibration: 2 g (10-500 Hz), 30 min for 2 sides
CommunicationUSB A, USB B, RJ45, WiFi, Bluetooth
LocalizationEnglish, Chinese, Japanese, Russian, German, French, Italian, and Spanish
Warranty1 year
DescriptionSpecifications
Readout Accuracy for 100 ohm PRT (Probe Accuracy Not Included)±0.009°C at -40°C
Readout Resolution1 mΩ
Reference Resistance Range0 Ω to 400 Ω
Reference Resistance Accuracy0 Ω to 50 Ω: 0.002 Ω
Reference CharacterizationsITS-90, CVD, IEC-751, Resistance
Reference Measurement Capability4-wire PRT
Reference Probe Connection6-pin lemo smart connector
RTD Channels2
RTD Measurement Accuracy (excl sensor) Compliance0 Ω to 25 Ω: 0.002 Ω
RTD Measurement Resolution0 Ω to 400 Ω: 1 mΩ
RTD Measurement Resistance Range0 Ω to 4K Ω
RTD CharacterizationsPT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120
RTD ConnectionFour 4 mm input jacks
RTD Channels2 channels. Both accept 2, 3, or 4-wire RTDs
TC Channel2
TC Measurement ChannelsMini TC terminals: Accepting S, R, K, B, N, E, J, T, C, D, G, L, and U
TC Measurement Accuracy (excl sensor)Type K: ±0.13°C at 0°C ±0.15°C at 155°C ±0.18°C at 350°C ±0.24°C at 660°C
TC Range–100 mV to 100 mV
TC Resolution0.001 mV, Input Impedance<1 MΩ
TC Voltage Accuracy0.02% RD + 5 µV
Internal CJC Accuracy±0.35°C (ambient from 0 °C to 50 °C)
Current Range–30 mA to 30 mA
Current Accuracy0.02% RD + 2 µA
Current Resolution0.001 mA, Input Impedance: < 10Ω
Voltage Range–30 V to 30 V
Voltage Accuracy±0.02% RD + 2 mV
Voltage Resolution0.001V; Input impedance: < 1MΩ
Switch TestMechanical or Electrical
DC 24V Output24 V ±1 V, MAX60 mA
Hart CommunicationOptional (ADT875PC Model)
DocumentationUp to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions.
Temperature Coefficient 0℃ to 8℃ and 38℃ to 50℃ADT875(PC)-155: ±0.005 °C/°C
TC TypeTemperature (°C)
B250
350±1.44
660±0.84
C0
155±0.34
350±0.33
660±0.38
D0
155±0.37
350±0.33
660±0.36
E-40
0±0.09
155±0.1
350±0.13
660±0.19
G0
155±0.71
350±0.43
660±0.36
J-40
0±0.1
155±0.12
350±0.16
660±0.21
K-40
0±0.13
155±0.16
350±0.19
660±0.25
     
  • ADT875-660-B-110V dry well calibrator
  •  
  • Powe Adapter
  •  
  • USB Cable
  •  
  • Insert removal tool 
  •  
  • Thermal Shield
  •  
  • Test Leads  2 Sets
  •  
  • Certificate of calibration
  •  
  • CD Manual