Fluke 1586A 120/C Super-DAQ Precision Temperature Scanner, Mainframe Only, 120V, Accredited Calibration Specs

Mains Voltage100 V Setting: 90 V to 110 V   120 V Setting: 108 V to 132 V   220 V Setting: 198 V to 242 V   240 V Setting: 216 V to 264 V
Frequency47 Hz to 440 Hz
Power Consumption36 VA peak (24 W average)
Environment   TemperatureOperating: 0 °C to 50 °C   Full accuracy: 18 °C to 28 °C   Storage: −20 °C to 70 °C   Warm-up: 1 hour to full accuracy specifications
Relative Humidity   (non-condensing)Operating: 0 °C to 30 °C <80 %;                    30 °C to 50 °C <50 %   Storage: −20 °C to 70 °C <95 %
AltitudeOperating: 2,000 m   Storage: 12,000 m
Vibration and ShockComplies with MIL-PRF-28800F Class 3
Input Protection50 V all functions, terminals and ranges
Math ChannelsNumber of channels: 20   Operations: sum, difference, multiply, divide, polynomial, power, square   root, reciprocal, exponential, logarithm, absolute value, average, maximum,   minimum
Triggersinterval, external (trigger input), alarm, remote (bus), manual, automated   test
MemoryScan data RAM: 75,000 readings with timestamp   Data/Setup flash memory: 20 MB
USB Host PortConnector type: Type A   Function: Memory   File system: FAT32   Memory capacity: 32 GB
USB Device PortConnector type: Type B   Class: Instrument   Function: Control and data transfer   Command protocol: SCPI
LANFunction: Control and data transfer   Network protocols: Ethernet 10/100, TCP/IP   Command protocol: SCPI
RS-232Connector: D-sub 9 pin (DE-9)   Baud rates: 1200, 2400, 4800, 9600, 19200, 38400   Function: Temperature source control output
DimensionsHeight: 150 mm   Width: 245 mm   Depth: 385 mm Weight: 6 kg (typical configuration)   Shipping Weight: 9.5 kg (typical configuration)
ConformityCE, CSA, IEC 61010 3rd ed.
Scan rateFast: 10 channels per second max (0.1 seconds per channel) Medium: 1 channel   per second (1 second per channel)   Slow: 4 seconds per channel
Display Resolution6 1/2 digits (see Measurement Characteristics tables below to find the display   resolution of temperature readings)
Temperature Range−200 °C to 1200 °C (depending on the sensor)
Resistance Range0 Ω to 4 kΩ
Offset Compensation0 Ω to 400 Ω, 4-wire: automatic current   reversal   400 Ω to 4000 Ω or 3-wire: none
Source Current Reversal Interval (0 Ω to 400 Ω   range)Fast sample rate: 2 ms   Medium sample rate: 250 ms   Slow sample rate: 250 ms
Maximum Lead Resistance (4-wire Ω)2.5 % of range per lead for 400 Ω and 4 kΩ   ranges
RangeSample Rate
0 Ω to   400 ΩSlow
Medium0.002 % or 0.002 Ω
Fast0.002 % or 0.005 Ω
400 Ω to   4 kΩSlow
Medium0.004 % or 0.1 Ω
Fast0.004 % or 0.18 Ω
Sample   RateTemperature
Slow−200 °C
0 °C0.005 °C
300 °C0.012 °C
600 °C0.02 °C
Medium−200 °C
0 °C0.005 °C
300 °C0.012 °C
600 °C0.02 °C
Fast−200 °C
0 °C0.013 °C
300 °C0.014 °C
600 °C0.02 °C
RangeSlow / Medium Sample Rate
0 Ω to 400 Ω0.001 °C
400 Ω to 4 kΩ0.001 °C
Temperature Range−200 °C to 400 °C (depending on the sensor)
Resistance Range0 Ω to 1 MΩ
RangeSlow Sample Rate
0 Ω to 2.2 kΩ0.004 % + 0.2 Ω
2.1 kΩ to 98 kΩ0.004 % + 0.5 Ω
95 kΩ to 1 MΩ0.015 % + 5 Ω
TemperatureSlow Sample Rate
−40 °C0.001 °C
0 °C0.003 °C
25 °C0.006 °C
50 °C0.008 °C
100 °C0.047 °C
150 °C0.23 °C
TemperatureSlow Sample Rate
−40 °C0.003 °C
0 °C0.002 °C
25 °C0.004 °C
50 °C0.005 °C
100 °C0.022 °C
150 °C0.096 °C
TemperatureSlow Sample Rate
−40 °C0.003 °C
0 °C0.002 °C
25 °C0.003 °C
50 °C0.005 °C
100 °C0.011 °C
150 °C0.04 °C
RangeSlow / Medium   Sample Rate
0 Ω to 2.2 Ω0.0001 °C
2.1 Ω to 98 Ω0.0001 °C
95 Ω to 1 MΩ0.0001 °C
Temperature Range−270 °C to 2315 °C (depending on the sensor)
Voltage Range−15 mV to 100 mV
RangeAccuracy Channel   1
−15 mV to 100 mV0.004 % + 4 μV
ModuleCJC Accuracy
DAQ-STAQ Module0.25 °C
High-Capacity Module0.6 °C