Associated Research 07850 – HypotULTRA, AC/DC Hipot with Insulation Resistance
$3,347.00
In Stock
In Stock
Associated Research 07850 - HypotULTRA, AC/DC Hipot with Insulation Resistance Specs
| Voltage | 100 – 120 VAC / 200 – 240 VAC ± 10% Auto Range |
| Frequency | 50/60 Hz ± 5% |
| Fuse | 7804/7820/7850: |
| 7800/7854: | 15A, Fast Blow 250 VAC |
| Output Voltage | Range: Resolution: Accuracy: |
| Output Frequency | 50/60 Hz ± 0.1%, User Selection |
| Output Waveform | Sine Wave, Crest Factor = 1.3 - 1.5 |
| Output Regulation | ± (1% of output + 5V) |
| HI and LO-Limit Total | Total |
| Real | Range: Resolution: Range: Resolution: Accuracy: |
| Ramp Up Timer | Range: |
| Ramp Down Timer | Range: |
| Dwell Timer | Range: |
| Ground Continuity | Current: DC 0.1A ± 0.01A, fixed |
| Current | Max. Ground Resistance: 1.0 Ω ± 0.1 Ω |
| Arc Detection | Range: |
| Output Voltage | Range: Resolution: Accuracy: |
| DC Output Ripple | <4% (6 KV/10 mA at Resistive Load) |
| HI and LO-Limit | Range: Resolution: Accuracy: |
| Range: Resolution: Accuracy: | 1.000 - 9.999 μA 0.001 μA ± (2% of setting + 10 counts), Low Range is ON |
| Range: Resolution: Accuracy: | 10.00 - 99.99 μA 0.01 μA ± (2% of setting + 10 counts), Low Range is ON |
| Range: Resolution: Accuracy: | 100.0 – 999.9 μA 0.1 μA ± (2% of setting + 2 counts) |
| Range: Resolution: Accuracy: | 1,000 - 20,000 μA range (7804/54) 1,000 - 10,000μA range (7800/50) 1 μA ± (2% of setting + 2 counts) |
| Ramp Up Timer | Range: |
| Ramp Down Timer | Range: |
| Dwell Timer | Range: |
| Ramp-HI Selectable | Range: |
| Charge-LO | Range: |
| Discharge Time | < 50 ms for no load, < 100 ms for capacitive load |
| Maximum Capacitive Load DC Mode | 1μF < 1kV 0.75 μF < 2 kV 0.5 μF < 3 kV |
| Arc Detection | Range: Resolution: Accuracy: |
| Output Voltage, DC | Range: Resolution: Accuracy: |
| Range: Resolution: Accuracy: | 1,001 - 6,000 VDC 1 VDC ± (2% of setting + 5 V) |
| Charging Current HI and LO-Limit | Maximum > 20 mA peak |
| Range: Resolution: Accuracy: | 0.10 MΩ – 99.9 MΩ (HI-Limit: 0=OFF) 0.01 MΩ ± (2% of setting + 2 counts) |
| Range: Resolution: Accuracy: | 100.0 MΩ – 999.9 MΩ 0.1 MΩ 1,000 – 9,999 ± (5% of setting + 2 counts) |
| Range: Resolution: Accuracy: | 1,000 MΩ – 50,000 MΩ 1 MΩ 10,000 – 50,000 ± (15% of setting + 2 counts) |
| Ramp Up Timer | Range: |
| Ramp Down Timer | Range: |
| Dwell Timer | Range: |
| Delay Timer | Range: |
| Charge-LO | 0.000 - 3.500 μA or Auto Set |
| Output Current, DC | 1 A for 0.000 - 1.000 Ω, 0.1 A for 1.01 - 10.00 Ω 0.01 A for 10.01 - 100 Ω, 0.001 A for 101 - 1,000 Ω 0.0001 A for 1001 - 10,000 Ω, 1 A is Max |
| Resistance Display Max & Min Max-Lmt | Range: Resolution: Accuracy: |
| Range: Resolution: Accuracy: | 1.01 - 10.00 Ω 0.01 Ω ± (1% of setting + 3 counts) |
| Range: Resolution: Accuracy: | 10.1 - 100.0 Ω 0.1 Ω ± (1% of setting + 3 counts) |
| Range: Resolution: Accuracy: | 101 - 1,000 Ω 1 Ω ± (1% of setting + 3 counts) |
| Range: Resolution: Accuracy: | 1,001 - 10,000 Ω 1 Ω ± (1% of setting + 10 counts) |
| Dwell Timer | Range: Resolution: Accuracy: |
| Resistance Offset | Range: Resolution: Accuracy: |
| Memory | 2,000 steps, 200 steps per test file max 100,000 test results |
| Mechanical | Bench or rackmount (2U height) with feet |
| Interface | Standard: USB, RS-232 Optional: GPIB (IEEE-488.2), Ethernet or USB Printer |
| SmartGFI® | 0, 0.4 - 5.0 mA (0=OFF) |
| Dimensions (W x H x D) | 16.92" x 3.50" x 15.75" (430 x 88.1 x 400mm) |
- NIST Traceable Calibration Certificate
- High Voltage Cable
- (2) Return Cable
- USB Cable AB-Type 1.8m
- Fuse
- Line Cord
- Adapter Box
- (2) 2U Rack Mount Handle
- (2) 2U Rack Mount Bracket
- (4) Screw M4 x 12mm FHMS
- Interlock Connector
- 8GB USB Flash Drive






Reviews