Vitrek 957i - 15KVDC 6KVAC/IR/LR Electrical Safety Compliance Analyzer Specs

DC Output Voltage20V to 6500V (951i & 952i)   50V to 11000V (953i, 954i & 955i)   100V to 15000V (957i)   DC Dielectric Tests (DCW, DCIR, DCez) : Accuracy: 0.25% + 0.5V (1 year 23°C ± 3°)
DC Current Sourcing50mA max, 25mA above 6000V (951i & 952i)   30mA max, 20mA above 6000V, 10mA above 7500V,   5mA above 9000V (953i, 954i, 955i) 10mA (957i)
Ramp Time0.01 to 9999sec, 0.01sec resolution or 0.1 to 50000V/sec, 0.1V/sec resolution
Dwell Time0.02 to 9999 seconds or user terminated, 0.01sec resolution
DC Leakage CurrentMeasurement Range: 0 to +/-200mA,   Resolution: 4 digits (9999 counts) down to 100pico-amps   Accuracy: 0.25% + 0.5nA + ½ digit (1 year 23°C ± 3°)   Selectable Min & Max limits for Ramp & Dwell, from 100 pico-amps up   Measurement Period: 1 power line cycle (50/60Hz)
Insulation Resistance (IR)Test modes include: End on pass reading, end on fail or end on timer
Measurement Range0 to 150Kohm (999.9mΩ to 99.99KΩ, 149.9KΩ in 7 ranges).
Resolution4 digits, 100µΩ on 1ohm range
Test Current55mAdc constant current up to ≈ 91Ω, 5VDC constant V above
Accuracy (4-wire)0.5% + 0.002ohm + ½ digit up to 30K ohm   1.5% + ½ digit above 30K ohm   5% + 1 digit from 100K to 150K ohm   Add 20mΩ for 2 -wire
AC Output Voltage20V to 6000V RMS (951i, 952i, 953i, 954i)   50V to 10,000V RMS (955i)   Resolution: 0.1V up to 999.9V, 1V above   Accuracy: 0.25% + 1.5V (+ 0.01% per Hz above 100Hz)   Decrease max output voltage by 0.1% per Hz above 100Hz   Decrease max voltage by 12.5V/mA loading (25V/ma 955i)
AC Current Sourcing50mA RMS max (951i, 952i, 953i, 954i)   30mA RMS max (955i)   100mA RMS max with 500VA option (951i, 953i)   200mA RMS with option AC-2 (2KVAC RMS max output)
Output FrequencyDigitally synthesized, low distortion sinewave   20Hz to 500Hz, standard, 500VA or AC-2)   40Hz to 500Hz, (955i)   0.1% accuracy, 0.1Hz resolution (1Hz above 99.9Hz)
Ramp Time0 to 9999sec, 0.01sec resolution or 0.1 to 100000V/sec, 0.1V/sec resolution
Dwell Time0.02 to 9999 seconds or user terminated, 0.01sec resolution
AC Leakage CurrentMeasurement Range: 0 to +/-200mA RMS   Resolution: 4 digits (9999 counts) down to 100pico-amps   Accuracy: 0.5% + 10nA (add 0.005% per Hz above 100Hz)   Selectable min & max limits for Ramp & Dwell, from 100 pico-amps up   Measurement Period: 1 power line cycle (50/60Hz)
Phase MeasurementTotal RMS current, In-phase current, Quadrature current (reactive/out-of-phase)   Accuracy: 0.01° per Hz, relative to output voltage
Voltage Measurement0 to ± 8KVDC 6KV RMS AC (951i & 952i)   0 to ± 11KVDC 8KV RMS AC (953i & 954i)   0 to ± 11KVDC 10KV RMS AC ( 955i)   Resolution: 0.1V, 1 V above 999.9V   DC Accuracy: 0.25% + 0.5V AC Accuracy: 0.5% + 1.5V
Leakage Current0 to ± 200mA DC or RMS AC   Resolution: 4 digits (9999 counts) down to 100pico-amps   DC Accuracy: 0.25% + 0.5nA AC Accuracy: 0.5% + 20nA
Test ResultsTest Time: 0.02 to 9999 sec   Last, Minimum, Average & Max V& A reading plus arc current
Test WaveformTrapezoidal (Selectable positive polarity, negative polarity or bi-polar)
Ramp Up/Down Time1ms (0.5ms for option AC2) to 30mS with 0.1ms resolution
Dwell Time1ms (0.5ms for option AC2) to 30mS with 0.1ms resolution
Test Voltage50V to 8000V (20V to 2750V with option AC-2)
Resolution0.1V up to 999.9V, 1V above
Accuracy0.25% + 1.5V
Arc DetectionTest Specific, Dual Parameter. Allows a specific broadband current amplitude limit from 2 to 20mA peak and pulse width limit from 4 to 30 microseconds for each test
EthernetHigh speed, high noise immunity LAN interface
RS232 InterfaceSelectable baud: 9600, 19200, 38400, 57600 or 115200, full handshake
VICL InterfacesTwo each provided for control of HV Scanners and other 950i series units
Digital I/O InterfaceProvides 8 digital inputs and 5 digital outputs. Functions include Test Selection, Start/Stop, Testing, Pass/Fail, Print, HV Present, Safety Interrupt, Dwell Timer
USB Host Printer PortFor hard copy test reports and LAN/Ethernet Interface
Optional GPIBOption GP-9 adds GPIB capability to LAN/Ethernet card
Test Lead Safety SenseTLSS™ Technology continuously verifies that test leads are properly connected prior to and during HV, 4-wire Low Ohms and Ground Bond testing
Real Time ClockAccuracy: 10 seconds per day, Battery Backup: 30 days minimum
Non-Volatile Memory100 user test sequences up to 100 steps each not to exceed 1000 total test steps. All test sequences, user settings and calibration data are stored in internal non-volatile Memory data retention is specified for 20 years and 1000000 write cycles
Dwell Time Accuracy0.05% + 20mS, Digital output provides dwell timer verification
Operating Temperature0 °C to 50 °C
Humidity90% RH max, 0 to 40 C
Power110 to 260 VAC, 50-60 Hz, 500VA Max
Dimensions89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D)
Weight9Kg (18 lb.) Net / 18Kg (25 lb.) shipping (951i, 953i, 959i)   14Kg (28 lb.) Net / 18Kg (35 lb.) shipping (952, 954i, 955i)
AccessoriesAlligator test leads, NIST traceable calibration certificate with no data, power cord and operator's manual. ISO 17025 cal cert with data and uncertainties available
WarrantyOne year parts and labor standard, 3 year extended warranty with registration and annual factory calibration
  • 15KVDC 6KVAC/IR/LR Electrical Safety Compliance Analyzer
  • TL-209
  • Power Cord
  • Operating Manual
  • QTPro II-950 45 day free trial
  • NIST Traceable Calibration Certificate without data

Reviews

There are no reviews yet.

Be the first to review “Vitrek 957i – 15KVDC 6KVAC/IR/LR Electrical Safety Compliance Analyzer”

Your email address will not be published. Required fields are marked *